Goldway  Bedside patient monitor

Goldway G30E

Bedside patient monitor

Najdi podobne izdelke

Clinicians need to respond swiftly to rapid changes in their patient's condition. Having a highly reliable, versatile patient monitoring solution on hand saves precious time in determining the best course of care. Philips Goldway G30E is your choice for a cost-effective, easy to learn, pre-configured bedside monitor that provides reliable measurements (Philips ST/AR ECG and FAST-SpO₂).

Features
Advanced functionalities

Advanced functionalities

To accommodate different clinical environments, the G30E delivers real-time data about cardiac and respiratory conditions. In addition to basic monitoring parameters, such as Philips ST/AR 3- and 5-lead ECG/ RESP, NBP, 2 temperatures, and Philips SpO₂/PR, ST and arrhythmia monitoring, it also provides 2-channel invasive blood pressure and microstream etCO₂ monitoring as options.
Ease of use

Ease of use

The monitor has a 10.4” display with flexible display modes, such as big number display mode, minitrend, OxyCRG, as well as auto waveform channel height adjustment function to fit various clinical settings. Intuitive menus, fixed key access, bright color screen and optional touchscreen were all designed for maximum ease of use.
Networking and data management

Networking and data management

The G30E comes standard with LAN networking capabilities for facilities looking to create a network with a bi-directional communication with Philips Efficia CMS200 central monitoring system or require data output in HL7 format.
Compact and cost-effective

Compact and cost-effective

Built according to Philips exacting quality standards, the G30E patient monitor offers market-proven essential parameters in a compact, cost-effective monitor to provide accurate measurement results within your tight budget. The alarm latch function helps reduce alarm fatigue to form a silent care environment.
Works with Efficia consumables

Works with Efficia consumables

The G30E patient monitor is designed to be compatible with Philips Efficia consumables, which are also used on Philips Efficia CM series bedside patient monitors, to lower your total cost of ownership by sharing Philips consumables across different care settings.
  • Advanced functionalities
  • Ease of use
  • Networking and data management
  • Compact and cost-effective
See all features
Advanced functionalities

Advanced functionalities

To accommodate different clinical environments, the G30E delivers real-time data about cardiac and respiratory conditions. In addition to basic monitoring parameters, such as Philips ST/AR 3- and 5-lead ECG/ RESP, NBP, 2 temperatures, and Philips SpO₂/PR, ST and arrhythmia monitoring, it also provides 2-channel invasive blood pressure and microstream etCO₂ monitoring as options.
Ease of use

Ease of use

The monitor has a 10.4” display with flexible display modes, such as big number display mode, minitrend, OxyCRG, as well as auto waveform channel height adjustment function to fit various clinical settings. Intuitive menus, fixed key access, bright color screen and optional touchscreen were all designed for maximum ease of use.
Networking and data management

Networking and data management

The G30E comes standard with LAN networking capabilities for facilities looking to create a network with a bi-directional communication with Philips Efficia CMS200 central monitoring system or require data output in HL7 format.
Compact and cost-effective

Compact and cost-effective

Built according to Philips exacting quality standards, the G30E patient monitor offers market-proven essential parameters in a compact, cost-effective monitor to provide accurate measurement results within your tight budget. The alarm latch function helps reduce alarm fatigue to form a silent care environment.
Works with Efficia consumables

Works with Efficia consumables

The G30E patient monitor is designed to be compatible with Philips Efficia consumables, which are also used on Philips Efficia CM series bedside patient monitors, to lower your total cost of ownership by sharing Philips consumables across different care settings.

Specifikacije

Safety approval and quality system
Safety approval and quality system
Safety standard
  • IEC 60601-1-2:2014
Quality system
  • ISO9001 & ISO13485 certified CE marking according to the Council Directive 93/42/EEC
Electromagnetic environment
  • IEC 60601-1-2:2000/A1:2014
Water proof
  • IPX1
Power supply
Power supply
Battery
  • Rechargeable Li-Ion battery Maximum 2.5 hour continuous monitoring (1 battery)
AC power input
  • 100 - 240 VAC, 50/60 Hz
ECG
ECG
ST measurement range
  • -2.5 mV - +2.5 mV
Waveform speed
  • 6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s
Lead selection, 3-lead
  • I, II, III
Lead type
  • 3-lead or 5-lead selectable
HR accuracy
  • ±5 bpm or ±1 %, whichever is greater
HR measurement
  • Adult / pediatric: 15 bpm - 300 bpm Neonatal: 15 bpm - 300 bpm Step: 1 bpm
Arrhythmia
  • 21 calls
Pacemaker detection
  • Yes
Protection
  • Against electrosurgical interference and defibrillation
Gain selection
  • x 1/4, x 1/2, x 1, x 2, x 4, Auto
Lead selection, 5-lead
  • I, II, III, aVR, aVL, aVF, V
Display
Display
Channels
  • Max 11 channels (auto waveform channel height adjustment)
Type
  • Color TFT LCD, 4:3
Resolution
  • 800 x 600 pixels
Size
  • 10.4+611:631"
Data review and storage
Data review and storage
Data review and storage
  • 1600- NBP data, 1200 hr trend graph, 120 min full disclosure ECG 1, 200 - alarm & arrhythmia events
SpO2 (Philips SpO2 technology)
SpO2 (Philips SpO2 technology)
Pulse rate range
  • 30 bpm - 300 bpm
Pulse rate accuracy
  • ±2 bpm or ±2%, whichever is greater
Waveform speed
  • 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Accuracy
  • 70 % - 100 %, depends on sensor; 0 - 69 %, unspecified
Perfusion index
  • Yes
Measurement range
  • 0 - 100 %
Weight and dimensions
Weight and dimensions
Net weight
  • Approx. 3.5 kg without battery
Dimensions (w x d x h)
  • 316 mm x 265 mm x 131 mm (12.5 in. x 10.5 in. x 5.2 in.)
Temperature (TEMP, 2-channel)
Temperature (TEMP, 2-channel)
Units
  • ℃ / ℉
Resolution
  • 0.1℃ (0.2℉)
Accuracy
  • ±0.1℃ (±0.2℉) (without sensor)
Minimum measuring time
  • Attachable skin surface probe: less than 10 minutes Esophageal/rectal probe: less than 2 minutes
Measurement range
  • 0 - 50℃ (32℉ - 122℉)
Invasive Blood Pressure
Invasive Blood Pressure
Accuracy
  • ±4 mmHg (0.53 kPa) or ±4 % of reading, whichever is greater
Invasive blood pressure (IBP, 2-channel, optional)
Invasive blood pressure (IBP, 2-channel, optional)
Parameters
  • SYS/DIA/MAP/Pulse Rate
Units
  • mmHg/kPa/cmH₂O
Pulse rate range
  • 30 bpm – 254 bpm
Resolution
  • 1 mmHg (0.13 kPa)
Sampling rate
  • 62.5 samples per second
Measurement range
  • -50 - 400 mmHg (-6.6 kPa - 53.3 kPa)
Transducer sensitivity
  • 5 uV/V/mmHg
Waveform speed
  • 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Pulse rate accuracy
  • ±2bpm or ±2%, whichever is greater
Respiration (RESP)
Respiration (RESP)
Waveform speed
  • 6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s Accuracy undefined
Apnea alarm
  • Yes, 10 - 60 s
Measurement method
  • Thoracic impedance
Gain selection
  • x 1/2, x 1, x 2
Measurement range
  • 0 rpm - 150 rpm
RESP lead
  • RA-LL
Accuracy
  • 0rpm – 120rpm: ±1 rpm; 121rpm – 150rpm: ±2 rpm
Work mode
  • Auto or manual (threshold)
Non-invasive blood pressure (NBP)
Non-invasive blood pressure (NBP)
MAP measurement range, neonate
  • 20 mmHg – 125mmHg (2.7 kPa – 16.7kPa)
Operating mode
  • Manual/Auto/STAT
MAP measurement range, adult / pediatric
  • 20 mmHg – 235 mmHg (2.7 kPa – 31.3 kPa)
Accuracy, maximum standard deviation
  • 8 mmHg (1.1 kPa)
Systolic measurement range, adult/ pediatric
  • 30 mmHg - 254 mmHg (4.0 kPa - 33.9 kPa)
Cuff pressure range, neonate
  • 0 - 150 mmHg ( 0 kPa - 20.0 kPa)
Cuff pressure range, adult / pediatric
  • 0 - 300 mmHg ( 0 kPa - 40.0 kPa)
Pulse rate range
  • 40 bpm – 240 bpm
Diastolic measurement range, neonate
  • 10 mmHg - 110 mmHg (1.3 kPa - 14.7 kPa)
Accuracy, mean error
  • ±5 mmHg (±0.7 kPa)
Pulse rate accuracy
  • ±5bpm or ±5%, whichever is greater
Method
  • Automatic oscillometric
Units
  • mmHg/kPa
Systolic measurement range, neonate
  • 30 mmHg - 135 mmHg (4.0 kPa - 18.0 kPa)
Diastolic measurement range, adult / pediatric
  • 10 mmHg - 220 mmHg (1.3 kPa - 29.3 kPa)
Parameters
  • SYS/DIA/MAP/Pulse Rate
Auto measure intervals
  • 3, 5, 10, 15, 30, 60, 90, 120, 240, 480 minutes
Capnopgraphy (etCO2, optional)
Capnopgraphy (etCO2, optional)
Measurement range
  • 0 - 99 mmHg (0 kPa - 13.2 kPa)
Method
  • Microstream, infrared absorption (Oridion)
awRR accuracy, 121 rpm – 150 rpm
  • ±3 rpm
Accuracy, 39 mmHg – 99 mmHg (5.2 kPa – 13.2 kPa)
  • ± (5% of reading + 0.08% for every 1 mmHg (0.13 kPa) above 38 mmHg (5.1kPa)
awRR accuracy, 71 rpm – 120 rpm
  • ±2 rpm
awRR accuracy, 3 rpm – 70 rpm
  • ±1 rpm
awRR range
  • 0-150 rpm
Parameters
  • awRR, etCO₂, inCO₂
Accuracy+C781, 0 – 38 mmHg (0 kPa – 5.1 kPa)
  • ±2 mmHg (±0.28 kPa)
Recorder (optional)
Recorder (optional)
Print speed
  • Auto, 12.5 mm/s, 25 mm/s, 50 mm/s
Type
  • Built-in thermal recorder, 3 channels
Print mode
  • Real-time recording, timed recording, alarm-activated recording
External interface (standard)
External interface (standard)
External interface
  • ECG analog output VGA output Nurse call LAN USB
Safety approval and quality system
Safety approval and quality system
Safety standard
  • IEC 60601-1-2:2014
Quality system
  • ISO9001 & ISO13485 certified CE marking according to the Council Directive 93/42/EEC
Power supply
Power supply
Battery
  • Rechargeable Li-Ion battery Maximum 2.5 hour continuous monitoring (1 battery)
AC power input
  • 100 - 240 VAC, 50/60 Hz
See all specifications
Safety approval and quality system
Safety approval and quality system
Safety standard
  • IEC 60601-1-2:2014
Quality system
  • ISO9001 & ISO13485 certified CE marking according to the Council Directive 93/42/EEC
Electromagnetic environment
  • IEC 60601-1-2:2000/A1:2014
Water proof
  • IPX1
Power supply
Power supply
Battery
  • Rechargeable Li-Ion battery Maximum 2.5 hour continuous monitoring (1 battery)
AC power input
  • 100 - 240 VAC, 50/60 Hz
ECG
ECG
ST measurement range
  • -2.5 mV - +2.5 mV
Waveform speed
  • 6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s
Lead selection, 3-lead
  • I, II, III
Lead type
  • 3-lead or 5-lead selectable
HR accuracy
  • ±5 bpm or ±1 %, whichever is greater
HR measurement
  • Adult / pediatric: 15 bpm - 300 bpm Neonatal: 15 bpm - 300 bpm Step: 1 bpm
Arrhythmia
  • 21 calls
Pacemaker detection
  • Yes
Protection
  • Against electrosurgical interference and defibrillation
Gain selection
  • x 1/4, x 1/2, x 1, x 2, x 4, Auto
Lead selection, 5-lead
  • I, II, III, aVR, aVL, aVF, V
Display
Display
Channels
  • Max 11 channels (auto waveform channel height adjustment)
Type
  • Color TFT LCD, 4:3
Resolution
  • 800 x 600 pixels
Size
  • 10.4+611:631"
Data review and storage
Data review and storage
Data review and storage
  • 1600- NBP data, 1200 hr trend graph, 120 min full disclosure ECG 1, 200 - alarm & arrhythmia events
SpO2 (Philips SpO2 technology)
SpO2 (Philips SpO2 technology)
Pulse rate range
  • 30 bpm - 300 bpm
Pulse rate accuracy
  • ±2 bpm or ±2%, whichever is greater
Waveform speed
  • 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Accuracy
  • 70 % - 100 %, depends on sensor; 0 - 69 %, unspecified
Perfusion index
  • Yes
Measurement range
  • 0 - 100 %
Weight and dimensions
Weight and dimensions
Net weight
  • Approx. 3.5 kg without battery
Dimensions (w x d x h)
  • 316 mm x 265 mm x 131 mm (12.5 in. x 10.5 in. x 5.2 in.)
Temperature (TEMP, 2-channel)
Temperature (TEMP, 2-channel)
Units
  • ℃ / ℉
Resolution
  • 0.1℃ (0.2℉)
Accuracy
  • ±0.1℃ (±0.2℉) (without sensor)
Minimum measuring time
  • Attachable skin surface probe: less than 10 minutes Esophageal/rectal probe: less than 2 minutes
Measurement range
  • 0 - 50℃ (32℉ - 122℉)
Invasive Blood Pressure
Invasive Blood Pressure
Accuracy
  • ±4 mmHg (0.53 kPa) or ±4 % of reading, whichever is greater
Invasive blood pressure (IBP, 2-channel, optional)
Invasive blood pressure (IBP, 2-channel, optional)
Parameters
  • SYS/DIA/MAP/Pulse Rate
Units
  • mmHg/kPa/cmH₂O
Pulse rate range
  • 30 bpm – 254 bpm
Resolution
  • 1 mmHg (0.13 kPa)
Sampling rate
  • 62.5 samples per second
Measurement range
  • -50 - 400 mmHg (-6.6 kPa - 53.3 kPa)
Transducer sensitivity
  • 5 uV/V/mmHg
Waveform speed
  • 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Pulse rate accuracy
  • ±2bpm or ±2%, whichever is greater
Respiration (RESP)
Respiration (RESP)
Waveform speed
  • 6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s Accuracy undefined
Apnea alarm
  • Yes, 10 - 60 s
Measurement method
  • Thoracic impedance
Gain selection
  • x 1/2, x 1, x 2
Measurement range
  • 0 rpm - 150 rpm
RESP lead
  • RA-LL
Accuracy
  • 0rpm – 120rpm: ±1 rpm; 121rpm – 150rpm: ±2 rpm
Work mode
  • Auto or manual (threshold)
Non-invasive blood pressure (NBP)
Non-invasive blood pressure (NBP)
MAP measurement range, neonate
  • 20 mmHg – 125mmHg (2.7 kPa – 16.7kPa)
Operating mode
  • Manual/Auto/STAT
MAP measurement range, adult / pediatric
  • 20 mmHg – 235 mmHg (2.7 kPa – 31.3 kPa)
Accuracy, maximum standard deviation
  • 8 mmHg (1.1 kPa)
Systolic measurement range, adult/ pediatric
  • 30 mmHg - 254 mmHg (4.0 kPa - 33.9 kPa)
Cuff pressure range, neonate
  • 0 - 150 mmHg ( 0 kPa - 20.0 kPa)
Cuff pressure range, adult / pediatric
  • 0 - 300 mmHg ( 0 kPa - 40.0 kPa)
Pulse rate range
  • 40 bpm – 240 bpm
Diastolic measurement range, neonate
  • 10 mmHg - 110 mmHg (1.3 kPa - 14.7 kPa)
Accuracy, mean error
  • ±5 mmHg (±0.7 kPa)
Pulse rate accuracy
  • ±5bpm or ±5%, whichever is greater
Method
  • Automatic oscillometric
Units
  • mmHg/kPa
Systolic measurement range, neonate
  • 30 mmHg - 135 mmHg (4.0 kPa - 18.0 kPa)
Diastolic measurement range, adult / pediatric
  • 10 mmHg - 220 mmHg (1.3 kPa - 29.3 kPa)
Parameters
  • SYS/DIA/MAP/Pulse Rate
Auto measure intervals
  • 3, 5, 10, 15, 30, 60, 90, 120, 240, 480 minutes
Capnopgraphy (etCO2, optional)
Capnopgraphy (etCO2, optional)
Measurement range
  • 0 - 99 mmHg (0 kPa - 13.2 kPa)
Method
  • Microstream, infrared absorption (Oridion)
awRR accuracy, 121 rpm – 150 rpm
  • ±3 rpm
Accuracy, 39 mmHg – 99 mmHg (5.2 kPa – 13.2 kPa)
  • ± (5% of reading + 0.08% for every 1 mmHg (0.13 kPa) above 38 mmHg (5.1kPa)
awRR accuracy, 71 rpm – 120 rpm
  • ±2 rpm
awRR accuracy, 3 rpm – 70 rpm
  • ±1 rpm
awRR range
  • 0-150 rpm
Parameters
  • awRR, etCO₂, inCO₂
Accuracy+C781, 0 – 38 mmHg (0 kPa – 5.1 kPa)
  • ±2 mmHg (±0.28 kPa)
Recorder (optional)
Recorder (optional)
Print speed
  • Auto, 12.5 mm/s, 25 mm/s, 50 mm/s
Type
  • Built-in thermal recorder, 3 channels
Print mode
  • Real-time recording, timed recording, alarm-activated recording
External interface (standard)
External interface (standard)
External interface
  • ECG analog output VGA output Nurse call LAN USB